Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters

Short peripheral catheters (SPC) are an existing conduit into many patients' veins and line draws from SPC are a desired method of routine blood collection especially in difficult venous access patients. The PIVO device facilitates blood collection through SPC and is being used clinically in a...

Full description

Saved in:
Bibliographic Details
Main Authors: Suzanne Adams (Author), Bridget Toroni (Author), Meenal Lele (Author)
Format: Book
Published: Hindawi Limited, 2018-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_1a131b8f7b0d4ed0a3db5881b70a3dd3
042 |a dc 
100 1 0 |a Suzanne Adams  |e author 
700 1 0 |a Bridget Toroni  |e author 
700 1 0 |a Meenal Lele  |e author 
245 0 0 |a Effect of the PIVO Device on the Procedure of Phlebotomy from Peripheral IV Catheters 
260 |b Hindawi Limited,   |c 2018-01-01T00:00:00Z. 
500 |a 2090-1429 
500 |a 2090-1437 
500 |a 10.1155/2018/7380527 
520 |a Short peripheral catheters (SPC) are an existing conduit into many patients' veins and line draws from SPC are a desired method of routine blood collection especially in difficult venous access patients. The PIVO device facilitates blood collection through SPC and is being used clinically in a number of hospitals. This study aimed to determine the appropriate wait time following a flush and the minimum waste volume required to obtain an undiluted blood sample when using the PIVO device and how that differed from current guidelines from SPC line draws. A clinical study was conducted examining the analyte results of samples drawn with PIVO through a SPC at varying wait times following a saline flush. Both an initial waste volume and a postwaste sample were compared to a venipuncture control. The resulting samples showed no saline dilution as measured by sodium and creatinine results at all studied wait times. These findings suggest that blood collections using the PIVO device can produce a clinically valid sample with a 30-second wait following a SPC flush and no waste volume prior to sample collection. 
546 |a EN 
690 |a Nursing 
690 |a RT1-120 
655 7 |a article  |2 local 
786 0 |n Nursing Research and Practice, Vol 2018 (2018) 
787 0 |n http://dx.doi.org/10.1155/2018/7380527 
787 0 |n https://doaj.org/toc/2090-1429 
787 0 |n https://doaj.org/toc/2090-1437 
856 4 1 |u https://doaj.org/article/1a131b8f7b0d4ed0a3db5881b70a3dd3  |z Connect to this object online.