Evaluation of the Haemoglobin Colour Scale and comparison with the HemoCue haemoglobin assay

OBJECTIVE: To evaluate the Haemoglobin Colour Scale developed by WHO for estimating haemoglobin concentration and to compare the results obtained using it and the HemoCue assay with those determined using a reference method, the Technicon H3 analyser. METHODS: The Colour Scale and HemoCue assay were...

Full description

Saved in:
Bibliographic Details
Main Author: Paddle J.J (Author)
Format: Book
Published: The World Health Organization, 2002-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_1f281526be3b45f087dc0cb4622a5a00
042 |a dc 
100 1 0 |a Paddle J.J.  |e author 
245 0 0 |a Evaluation of the Haemoglobin Colour Scale and comparison with the HemoCue haemoglobin assay 
260 |b The World Health Organization,   |c 2002-01-01T00:00:00Z. 
500 |a 0042-9686 
520 |a OBJECTIVE: To evaluate the Haemoglobin Colour Scale developed by WHO for estimating haemoglobin concentration and to compare the results obtained using it and the HemoCue assay with those determined using a reference method, the Technicon H3 analyser. METHODS: The Colour Scale and HemoCue assay were used to test 408 blood samples. Subsequently, Bland-Altman plots were determined and the proximity of the test results to those obtained using the reference method was determined. FINDINGS: The mean difference between the Haemoglobin Colour Scale and the reference method was 0.19 g/dl (95% confidence interval: 3.50 g/dl below to 3.11 g/dl above); the corresponding value for the HemoCue assay was 0.50 g/dl (1.16 g/dl below to 0.16 g/dl above). Only 46.08% of the results obtained by means of the Colour Scale were within 1.0 g/dl of the reference method, whereas 95.34% of the HemoCue results fell within this limit; 22.79% of the Colour Scale results but none of the HemoCue results lay more than 2.0 g/dl from the reference method. CONCLUSION: The Haemoglobin Colour Scale test is too inaccurate for general use, particularly if devices such as the HemoCue are available. 
546 |a EN 
690 |a Hemoglobinometry/methods 
690 |a Reference standards 
690 |a Comparative study 
690 |a Developing countries 
690 |a Public aspects of medicine 
690 |a RA1-1270 
655 7 |a article  |2 local 
786 0 |n Bulletin of the World Health Organization, Vol 80, Iss 10, Pp 813-816 (2002) 
787 0 |n http://www.scielosp.org/scielo.php?script=sci_arttext&pid=S0042-96862002001000010 
787 0 |n https://doaj.org/toc/0042-9686 
856 4 1 |u https://doaj.org/article/1f281526be3b45f087dc0cb4622a5a00  |z Connect to this object online.