Effects of clinical use of NiTi reciprocating instruments on cyclic and torsional resistance, and on roughness

Abstrac The aim of the present study was to evaluate the cyclic fatigue, torsional resistance and surface roughness of Reciproc R25 instruments in four different situations, namely as new instruments and as instruments tested after clinical preparation of one, two or three maxillary molars with four...

Full description

Saved in:
Bibliographic Details
Main Authors: Edson Jorge Lima MOREIRA (Author), Henrique dos Santos ANTUNES (Author), Victor Talarico Leal VIEIRA (Author), Daniele Moreira CAVALCANTE (Author), Henrique Eduardo OLIVEIRA (Author), Diogo da Silva OLIVEIRA (Author), Gustavo DE-DEUS (Author), Emmanuel João Nogueira Leal da SILVA (Author)
Format: Book
Published: Sociedade Brasileira de Pesquisa Odontológica, 2021-02-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_27f6d0f601dd4f29b0f39ce034d0b6d8
042 |a dc 
100 1 0 |a Edson Jorge Lima MOREIRA  |e author 
700 1 0 |a Henrique dos Santos ANTUNES  |e author 
700 1 0 |a Victor Talarico Leal VIEIRA  |e author 
700 1 0 |a Daniele Moreira CAVALCANTE  |e author 
700 1 0 |a Henrique Eduardo OLIVEIRA  |e author 
700 1 0 |a Diogo da Silva OLIVEIRA  |e author 
700 1 0 |a Gustavo DE-DEUS  |e author 
700 1 0 |a Emmanuel João Nogueira Leal da SILVA  |e author 
245 0 0 |a Effects of clinical use of NiTi reciprocating instruments on cyclic and torsional resistance, and on roughness 
260 |b Sociedade Brasileira de Pesquisa Odontológica,   |c 2021-02-01T00:00:00Z. 
500 |a 1807-3107 
500 |a 10.1590/1807-3107bor-2021.vol35.0021 
520 |a Abstrac The aim of the present study was to evaluate the cyclic fatigue, torsional resistance and surface roughness of Reciproc R25 instruments in four different situations, namely as new instruments and as instruments tested after clinical preparation of one, two or three maxillary molars with four root canals. The total time required to perform each root canal preparation was recorded. Cyclic fatigue resistance was determined by the time to fracture using a customized testing device (n = 10 per group). The torsional test evaluated the torque and angle of rotation to failure according to ISO 3630-1 (n = 10 per group). The roughness of the working parts of new and used instruments was evaluated with a profilometer (n = 5 per group). Statistical analysis was performed using one-way ANOVA and Tukey's test. The level of significance was set at 5%. No fractures or deformations were observed after clinical use. Higher preparation time was needed during the third use of the instruments for all root canals (p < 0.05). There were no significant differences among the groups in regard to either cyclic fatigue or torsional resistance (p > 0.05). Regarding the roughness measurements, groove depth was higher on new and one- versus two- or three-maxillary-molar-prepared instruments (p < 0.05). It can be concluded that the clinical use of Reciproc instruments increased preparation time and decreased surface roughness. However, clinical use did not affect the cyclic fatigue or torsional resistance of the Reciproc instruments. 
546 |a EN 
690 |a Dental Instruments 
690 |a Endodontics 
690 |a Root Canal Therapy 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Brazilian Oral Research, Vol 35 (2021) 
787 0 |n http://www.scielo.br/pdf/bor/v35/1807-3107-bor-35-e021.pdf 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1806-83242021000100218&tlng=en 
787 0 |n https://doaj.org/toc/1807-3107 
856 4 1 |u https://doaj.org/article/27f6d0f601dd4f29b0f39ce034d0b6d8  |z Connect to this object online.