Effect of implantoplasty on the elastic limit of dental implants of different diameters

Abstract Background Implantoplasty reduces both implant diameter and the thickness of its walls, subsequently reducing the ability of the implant to resist fracture in response to functional load. In combination with an increase in the crown-implant ratio due to bone loss, this could increase the le...

Full description

Saved in:
Bibliographic Details
Main Authors: Markel Diéguez-Pereira (Author), David Chávarri-Prado (Author), Iratxe Viteri-Agustín (Author), Oier Montalban-Vadillo (Author), Esteban Pérez-Pevida (Author), Aritza Brizuela-Velasco (Author)
Format: Book
Published: SpringerOpen, 2021-08-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_2a80ea655d58427d82330d16a2e94d05
042 |a dc 
100 1 0 |a Markel Diéguez-Pereira  |e author 
700 1 0 |a David Chávarri-Prado  |e author 
700 1 0 |a Iratxe Viteri-Agustín  |e author 
700 1 0 |a Oier Montalban-Vadillo  |e author 
700 1 0 |a Esteban Pérez-Pevida  |e author 
700 1 0 |a Aritza Brizuela-Velasco  |e author 
245 0 0 |a Effect of implantoplasty on the elastic limit of dental implants of different diameters 
260 |b SpringerOpen,   |c 2021-08-01T00:00:00Z. 
500 |a 10.1186/s40729-021-00363-6 
500 |a 2198-4034 
520 |a Abstract Background Implantoplasty reduces both implant diameter and the thickness of its walls, subsequently reducing the ability of the implant to resist fracture in response to functional load. In combination with an increase in the crown-implant ratio due to bone loss, this could increase the lever effect, which in presence of high masticatory forces or parafunctional habits, could lead to complications such as fracture of the implant or loosening of the prosthetic screw. Objectives To determine the elastic limits of internal connection, dental implants of different designs and diameters after an implantoplasty. Materials and methods This in vitro study included 315 tapered internal connection titanium dental implants, the threads of which were removed with an industrial milling machine-for standardized implantoplasty (IMP1; n = 105)-or with the conventional approach-manually, using high-speed burs (IMP2; n = 105). The remaining 105 implants were used as controls. The final implant diameters were recorded. The quality of the newly polished surfaces was assessed by scanning electron microscopy. All implants were subjected to a mechanical pressure resistance test. A Tukey's test for multiple comparisons was used to detect differences in the elastic limit and final implant diameters between the implant groups. Results There were statistically significant differences in the elastic limit between the IMP1, IMP2, and control groups (p < 0.05). Furthermore, the implant diameter was significantly smaller in the IMP1 and IMP2 groups (p < 0.05). Scanning electron microscopy revealed smooth implant surfaces in the IMP1 and IMP2 groups, with some titanium particles visible in the IMP1 group. Conclusions Implantoplasty significantly decreased the elastic limit of internal connection titanium dental implants, especially in those with a smaller diameter (3-3.5 mm). 
546 |a EN 
690 |a Medicine 
690 |a R 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n International Journal of Implant Dentistry, Vol 7, Iss 1, Pp 1-10 (2021) 
787 0 |n https://doi.org/10.1186/s40729-021-00363-6 
787 0 |n https://doaj.org/toc/2198-4034 
856 4 1 |u https://doaj.org/article/2a80ea655d58427d82330d16a2e94d05  |z Connect to this object online.