Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions
Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This re...
Saved in:
Main Authors: | , , |
---|---|
Format: | Book |
Published: |
Wolters Kluwer Medknow Publications,
2016-01-01T00:00:00Z.
|
Subjects: | |
Online Access: | Connect to this object online. |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This report highlights additional tips in transcatheter device retrieval. |
---|---|
Item Description: | 0974-2069 10.4103/0974-2069.180672 |