Successful retrieval of a Figulla Occlutech septal occluder - embolized device stability and potential solutions

Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This re...

Full description

Saved in:
Bibliographic Details
Main Authors: Masood Sadiq (Author), Asif Ur Rehman (Author), Amjad Mehmood (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2016-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Go to: We report a case of an atrial septal defect where a Figulla Occlutech device embolized into the right ventricle. As this device has no left atrial hub, we brought the bare device into the inferior vena cava and used a novel two-snare technique to slenderize the device into the sheath. This report highlights additional tips in transcatheter device retrieval.