Reconditioning of Debonded Pure Titanium Bracket (Using Micro-etcher)

Aims: To evaluate the degree of change that may occur in reconditioning the titanium bracket via micro-etcher. Materials and Methods: The sample consisted of (10) pure titanium brackets for each ofcontrol and reconditioning brackets. The parameters of the reconditioned bracket (slot width, slotdepth...

Full description

Saved in:
Bibliographic Details
Main Authors: Hussain A Obaidi (Author), Amer A Taqa (Author), Omar H AL-Luazy (Author)
Format: Book
Published: University of Mosul, College of Dentistry, 2008-04-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Aims: To evaluate the degree of change that may occur in reconditioning the titanium bracket via micro-etcher. Materials and Methods: The sample consisted of (10) pure titanium brackets for each ofcontrol and reconditioning brackets. The parameters of the reconditioned bracket (slot width, slotdepth, inter-wing gap, labio-lingual angle and base curvature angle) were measured and comparedwith that of the control bracket. Results: The results showed no significant difference between the controland reconditioned brackets of the slot parameters (width, depth, inter-win gap, labio-lingual angle)and the bracket's base curvature angle. While, the tensile bond strength of the control bracketsshowed significantly higher mean value as compared with the recycled group. Conclusions: The reconditioningof pure titanium brackets is recommended for reuse in the orthodontic treatment after sterilization.
Item Description:10.33899/rden.2008.9040
1812-1217
1998-0345