Experience with the largest custom-made 48 mm fenestrated atrial septal occluder device and its 4-year follow-up

Atrial septal defects (ASDs) measuring <38 mm are referred for transcatheter closure. Availability of larger devices up to 46 mm extended the inclusion criteria. An elderly hypertensive male with a 44 mm secundum ASD and coexistent sick sinus syndrome and atrioventricular (AV) nodal block present...

Full description

Saved in:
Bibliographic Details
Main Authors: Kothandam Sivakumar (Author), Gaurav Jain (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2023-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Atrial septal defects (ASDs) measuring <38 mm are referred for transcatheter closure. Availability of larger devices up to 46 mm extended the inclusion criteria. An elderly hypertensive male with a 44 mm secundum ASD and coexistent sick sinus syndrome and atrioventricular (AV) nodal block presented with syncope. Balloon interrogation unmasked restrictive left ventricular (LV) physiology. After AV synchronous pacing, balloon-assisted deployment of a custom fenestrated 48 mm Figulla septal occluder (Occlutech Inc., Schaffhausen, Switzerland) prevented a rise of LV end-diastolic pressures beyond 12 mmHg. Echocardiogram and computed tomography after 4 years confirmed a patent fenestration and favorable remodeling. This report of the clinical use of the largest ASD device demonstrated the feasibility of closure of extremely large defects despite a restrictive LV.
Item Description:0974-2069
10.4103/apc.apc_126_22