Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls
Objectives: To investigate the accuracy and reliability of implant site measurements, recorded from low-dose cone beam computed tomography (CBCT) images. Methods: CBCT reformatted images of five skulls were obtained using 40, 20 and 7 s exposure protocols. From these protocols, edentulous ridge dime...
Saved in:
Main Author: | |
---|---|
Format: | Book |
Published: |
Elsevier,
2012-07-01T00:00:00Z.
|
Subjects: | |
Online Access: | Connect to this object online. |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objectives: To investigate the accuracy and reliability of implant site measurements, recorded from low-dose cone beam computed tomography (CBCT) images. Methods: CBCT reformatted images of five skulls were obtained using 40, 20 and 7 s exposure protocols. From these protocols, edentulous ridge dimensions were recorded by two observers and compared with measurements recorded directly from the bone. The measurement errors and intra- and inter-examiner reliability were calculated for each exposure protocol and compared with each other. Results: The mean absolute errors from the 40, 20 and 7 s protocols were 0.50, 0.46, and 0.51 mm, respectively. The intra-examiner reliability scores were 0.996, 0.995 and 0.998, respectively. The inter-examiner reliability scores were 0.993, 0.998 and 0.994, respectively. There was no significant difference in accuracy or reliability between the three protocols. Conclusions: In imaging of dry skulls, lowering the CBCT exposure time from 40 s to 20 or 7 s does not affect the reliability or accuracy of implant site measurements. Keywords: CBCT, Radiation dosage, Reliability, Dental implantation, Dry skull |
---|---|
Item Description: | 1013-9052 10.1016/j.sdentj.2012.05.001 |