Assessing Exposures to Magnetic Resonance Imaging's Complex Mixture of Magnetic Fields for In Vivo, In Vitro, and Epidemiologic Studies of Health Effects for Staff and Patients

A complex mixture of electromagnetic fields is used in magnetic resonance imaging (MRI): static, low-frequency, and radio frequency magnetic fields. Commonly, the static magnetic field ranges from one to three Tesla. The low-frequency field can reach several millitesla and with a time derivative of...

Full description

Saved in:
Bibliographic Details
Main Authors: Jennifer Frankel (Author), Jonna Wilén (Author), Kjell Hansson Mild (Author)
Format: Book
Published: Frontiers Media S.A., 2018-03-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_4ae982c2cc794e5faa97817f30e76589
042 |a dc 
100 1 0 |a Jennifer Frankel  |e author 
700 1 0 |a Jonna Wilén  |e author 
700 1 0 |a Kjell Hansson Mild  |e author 
245 0 0 |a Assessing Exposures to Magnetic Resonance Imaging's Complex Mixture of Magnetic Fields for In Vivo, In Vitro, and Epidemiologic Studies of Health Effects for Staff and Patients 
260 |b Frontiers Media S.A.,   |c 2018-03-01T00:00:00Z. 
500 |a 2296-2565 
500 |a 10.3389/fpubh.2018.00066 
520 |a A complex mixture of electromagnetic fields is used in magnetic resonance imaging (MRI): static, low-frequency, and radio frequency magnetic fields. Commonly, the static magnetic field ranges from one to three Tesla. The low-frequency field can reach several millitesla and with a time derivative of the order of some Tesla per second. The radiofrequency (RF) field has a magnitude in the microtesla range giving rise to specific absorption rate values of a few Watts per kilogram. Very little attention has been paid to the case where there is a combined exposure to several different fields at the same time. Some studies have shown genotoxic effects in cells after exposure to an MRI scan while others have not demonstrated any effects. A typical MRI exam includes muliple imaging sequences of varying length and intensity, to produce different types of images. Each sequence is designed with a particular purpose in mind, so one sequence can, for example, be optimized for clearly showing fat water contrast, while another is optimized for high-resolution detail. It is of the utmost importance that future experimental studies give a thorough description of the exposure they are using, and not just a statement such as "An ordinary MRI sequence was used." Even if the sequence is specified, it can differ substantially between manufacturers on, e.g., RF pulse height, width, and duty cycle. In the latest SCENIHR opinion, it is stated that there is very little information regarding the health effects of occupational exposure to MRI fields, and long-term prospective or retrospective cohort studies on workers are recommended as a high priority. They also state that MRI is increasingly used in pediatric diagnostic imaging, and a cohort study into the effects of MRI exposure on children is recommended as a high priority. For the exposure assessment in epidemiological studies, there is a clear difference between patients and staff and further work is needed on this. Studies that explore the possible differences between MRI scan sequences and compare them in terms of exposure level are warranted. 
546 |a EN 
690 |a electromagnetic field 
690 |a occupational exposure 
690 |a switched gradient field 
690 |a diagnostic imaging 
690 |a measurement 
690 |a Public aspects of medicine 
690 |a RA1-1270 
655 7 |a article  |2 local 
786 0 |n Frontiers in Public Health, Vol 6 (2018) 
787 0 |n http://journal.frontiersin.org/article/10.3389/fpubh.2018.00066/full 
787 0 |n https://doaj.org/toc/2296-2565 
856 4 1 |u https://doaj.org/article/4ae982c2cc794e5faa97817f30e76589  |z Connect to this object online.