Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization

NiTi instruments have a high risk of separation due to torsional or flexural fatigue (cyclic fatigue). Chemomechanical preparation, cleaning procedures, chemical disinfection and sterilization cause the corrosion of endodontic instruments that may weaken the fracture resistance of the instruments. O...

Täydet tiedot

Tallennettuna:
Bibliografiset tiedot
Päätekijät: Ureyen Kaya BULEM (Tekijä), Ayse Diljin KECECI (Tekijä), Hilmi Egemen GULDAS (Tekijä)
Aineistotyyppi: Kirja
Julkaistu: University of São Paulo, 2013-12-01T00:00:00Z.
Aiheet:
Linkit:Connect to this object online.
Tagit: Lisää tagi
Ei tageja, Lisää ensimmäinen tagi!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_4ce36b2835834449a7f5e5e41e27e3d0
042 |a dc 
100 1 0 |a Ureyen Kaya BULEM  |e author 
700 1 0 |a Ayse Diljin KECECI  |e author 
700 1 0 |a Hilmi Egemen GULDAS  |e author 
245 0 0 |a Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization 
260 |b University of São Paulo,   |c 2013-12-01T00:00:00Z. 
500 |a 1678-7765 
500 |a 10.1590/1679-775720130083 
520 |a NiTi instruments have a high risk of separation due to torsional or flexural fatigue (cyclic fatigue). Chemomechanical preparation, cleaning procedures, chemical disinfection and sterilization cause the corrosion of endodontic instruments that may weaken the fracture resistance of the instruments. Objective To assess the effects of NaOCl immersion and autoclave sterilization on the cyclic fatigue resistance of ProFile, FlexMaster, Mtwo and TwistedFiles NiTi instruments (tip size 25, 0.06 taper, n=160). Material and Methods The instruments (n=10 for each subgroup) were dynamically immersed in NaOCl; immersed in NaOCl and sterilized in one autoclave cycle; 5 cycles immersed in NaOCl and sterilized in autoclave and not immersed in NaOCl and not sterilized (control group). Dynamic cyclic fatigue resistance was tested. The number of cycles to failure (NCF) were statistically analyzed (P<0.05). Results Kruskall-Wallis test indicated significant differences among the tested instruments in terms of NCF (P=0.000). The mean NCF of Mtwo (556.75) was higher than that of the Twisted File, Flexmaster and ProFile, 483.1, 376.12, 365.25, respectively. NaOCl immersion and autoclave sterilization have no effect on the NCF values of the tested instruments (P>.05). Conclusions Cyclic fatigue resistance of the tested NiTi instruments cannot be adversely affected by NaOCl immersion and autoclave sterilization. Production process (TwistedFiles) or design (Twisted Files, FlexMaster, Mtwo and ProFile) of the instruments can influence their cyclic fatigue resistance. 
546 |a EN 
690 |a Endodontics 
690 |a Fractures 
690 |a Sterilization 
690 |a Stress 
690 |a Sodium chloride 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Applied Oral Science, Vol 21, Iss 6, Pp 505-510 (2013) 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-77572013000600505&lng=en&tlng=en 
787 0 |n https://doaj.org/toc/1678-7765 
856 4 1 |u https://doaj.org/article/4ce36b2835834449a7f5e5e41e27e3d0  |z Connect to this object online.