Abutment Screw Loosening in Single Implant Restorations: Evaluation of Fracture Strength of Implant-Supported Cement-Retained Monolithic Zirconia Restorations Repaired with Porcelain and Composite Resin

Aim: To assess the fracture strength of repaired implant-supported cement-retained monolithic zirconia restorations with porcelain and composite resin. Materials and methods: Nighty monolithic zirconia crowns were cemented over the implant abutments, and divided into three groups (n = 30): control,...

Full description

Saved in:
Bibliographic Details
Main Authors: Hamid Neshandar Asli (Author), Mehran Falahchai (Author), Samiye Rahimabadi (Author), Hamidreza Arbab (Author), Zoheir Mousavi Mehr (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2021-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_53c4f30f2c0b475788a1c0cf5cfee332
042 |a dc 
100 1 0 |a Hamid Neshandar Asli  |e author 
700 1 0 |a Mehran Falahchai  |e author 
700 1 0 |a Samiye Rahimabadi  |e author 
700 1 0 |a Hamidreza Arbab  |e author 
700 1 0 |a Zoheir Mousavi Mehr  |e author 
245 0 0 |a Abutment Screw Loosening in Single Implant Restorations: Evaluation of Fracture Strength of Implant-Supported Cement-Retained Monolithic Zirconia Restorations Repaired with Porcelain and Composite Resin 
260 |b Wolters Kluwer Medknow Publications,   |c 2021-01-01T00:00:00Z. 
500 |a 2155-8213 
500 |a 10.4103/denthyp.denthyp_125_21 
520 |a Aim: To assess the fracture strength of repaired implant-supported cement-retained monolithic zirconia restorations with porcelain and composite resin. Materials and methods: Nighty monolithic zirconia crowns were cemented over the implant abutments, and divided into three groups (n = 30): control, screw access hole preparation in the crown and its restoration with composite resin (HC), and screw access hole preparation in the crown and its restoration with feldspathic porcelain (HP). Fracture strength was measured by a universal testing machine. Data were analyzed (α < 0.05). Results: The maximum fracture strength was in the control group (764.23 ± 108.41 N), whereas the minimum was in HC group (686.25 ± 101.86 N). The mean fracture strength in HC group was significantly lower than that in the control group (P = 0.008). No significant difference was noted between other groups (P > 0.05). Conclusion: Repaired implant-supported monolithic zirconia crowns with feldspathic porcelain provided a fracture strength comparable to that in the control group. 
546 |a EN 
690 |a implant supported dental prosthesis 
690 |a single-tooth implants 
690 |a y-tzp ceramic 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Dental Hypotheses, Vol 12, Iss 3, Pp 128-131 (2021) 
787 0 |n http://www.dentalhypotheses.com/article.asp?issn=2155-8213;year=2021;volume=12;issue=3;spage=128;epage=131;aulast=Asli 
787 0 |n https://doaj.org/toc/2155-8213 
856 4 1 |u https://doaj.org/article/53c4f30f2c0b475788a1c0cf5cfee332  |z Connect to this object online.