A simple protocol for the routine calibration of pH meters.

A simplified laboratory protocol for the calibration of pH meters is described and tested. It is based on the use of two analytical primary buffer solutions, potassium hydrogen phthalate and Borax (sodium tetraborate decahydrate) of precisely known concentrations and pH. The solutions may be stored...

Full description

Saved in:
Bibliographic Details
Main Authors: A. FEDERMAN NETO (Author), A. D.L. BORGES (Author), M. A.S. LAVRADOR (Author)
Format: Book
Published: São Paulo State University (UNESP), 2006-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A simplified laboratory protocol for the calibration of pH meters is described and tested. It is based on the use of two analytical primary buffer solutions, potassium hydrogen phthalate and Borax (sodium tetraborate decahydrate) of precisely known concentrations and pH. The solutions may be stored at room temperature for long periods, without decomposition and used directly. The calibration of the meter can be checked with standard solutions of sodium dihydrogen phosphate, sodium carbonate, sodium benzoate, sodium salicylate or potassium oxalate. Methods for the purification of Borax and potassium chloride are also given, and a new method for the neutralization of 0.9% saline is suggested.
Item Description:1808-4532
2179-443X