The influence of different screw tightening forces on the vertical misfit of implant-supported frameworks Avaliação do efeito da força de aperto de parafusos de retenção na desadaptação marginal de próteses sobre implantes

OBJECTIVES: The present in vitro study was designed to compare the differences in the vertical misfit of implant-supported frameworks using three different forces for tightening the bridge locking screws: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. METHODS:...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Diego Klee de Vasconcellos (Autor), Marco Antonio Bottino (Autor), Renato Sussumu Nishioka (Autor), Luiz Felipe Valandro (Autor), Elza Maria Valadares da Costa (Autor)
Formato: Libro
Publicado: University of São Paulo, 2005-06-01T00:00:00Z.
Materias:
Acceso en línea:Connect to this object online.
Etiquetas: Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_60e4e6fa3f0147fc8b3ab70bf464c073
042 |a dc 
100 1 0 |a Diego Klee de Vasconcellos  |e author 
700 1 0 |a Marco Antonio Bottino  |e author 
700 1 0 |a Renato Sussumu Nishioka  |e author 
700 1 0 |a Luiz Felipe Valandro  |e author 
700 1 0 |a Elza Maria Valadares da Costa  |e author 
245 0 0 |a The influence of different screw tightening forces on the vertical misfit of implant-supported frameworks Avaliação do efeito da força de aperto de parafusos de retenção na desadaptação marginal de próteses sobre implantes 
260 |b University of São Paulo,   |c 2005-06-01T00:00:00Z. 
500 |a 10.1590/S1678-77572005000200005 
500 |a 1678-7757 
500 |a 1678-7765 
520 |a OBJECTIVES: The present in vitro study was designed to compare the differences in the vertical misfit of implant-supported frameworks using three different forces for tightening the bridge locking screws: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. METHODS: The investigation was conducted based on the results given by 9 six-unit nickel-chromium (2 abutments/ 4 pontics) screw-retained implant-supported frameworks. The structures were exposed to simulated porcelain firings. The marginal misfit measurements were made using a traveling measuring microscope at selected screw tightening forces: fastening by hand until first resistance, and using torque drivers with 10 and 20Ncm. The results were submitted to one-way ANOVA with repeated measures on one factor, and post hoc pairwise comparisons using Tukey test (5%). RESULTS: The mean marginal misfit of the frameworks, fastening the screws by hand until first resistance, was 41.56µm (SD±12.45µm). The use of torque driver devices caused a significant reduction in marginal opening (p<0.05). With the lowest torque available (10Ncm), the mean marginal discrepancy at the abutment-framework interface was reduced an average of 52% to a mean marginal opening of 19.71µm (SD±2.97µm). After the use of the 20Ncm torque driver, the mean marginal discrepancy of the frameworks was reduced an average of 69% to a mean marginal opening of 12.82µm (SD±4.0µm). Comparing the use of torque drivers with 10 and 20 Ncm torque, the means are not significantly different from one another. CONCLUSION: The seating force has an important effect on the vertical misfit measurements, once it may considerably narrow the vertical misfit gaps at the abutment-framework interface, thus leading to a misjudgment of the real marginal situation.<br>O presente estudo avaliou o efeito da força de aperto de parafusos de retenção na desadaptação marginal de próteses sobre implantes. A investigação foi conduzida baseada em resultados fornecidos por nove estruturas metálicas implanto-suportadas de seis elementos (02 retentores e 04 pônticos), retidas por parafusos. A mensuração da desadaptação marginal foi realizada com auxílio de microscópio óptico, nos seguintes momentos: M1: aperto com chave manual até a primeira resistência do parafuso; M2: aperto utilizando torquímetro mecânico com 10Ncm e M3: aperto utilizando torquímetro mecânico com 20Ncm. As médias de desadaptação foram submetidas à análise estatística (ANOVA e Tukey). O emprego de torquímetros mecânicos, independente do valor de torque, determinou redução significativa na abertura marginal (p<0,05). A desadaptação média na interface conexão/prótese em M1 foi de 41,56µm (DP±12,45µm). O emprego do torquímetro de 10Ncm (M2), reduziu 52% desta fenda marginal, para uma média de desadaptação de 19,71µm (DP±2,97µm); enquanto o torquímetro de 20Ncm (M3) reduziu 69% da desadaptação marginal, para uma média de 12,82µm (DP±4,0µm) os de retenção de próteses implanto-suportadas tem importante efeito na mensuração da desadaptação marginal, uma vez que pode modificar consideravelmente as aberturas na interface conexão/prótese. 
546 |a EN 
690 |a Implantes dentários 
690 |a Osseointegração 
690 |a Prótese dentária fixada por implante 
690 |a Torque 
690 |a desadaptação 
690 |a Dental implants 
690 |a Osseointegration 
690 |a Implant-supported dental prosthesis 
690 |a Torque 
690 |a misfit 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Applied Oral Science, Vol 13, Iss 2, Pp 120-125 (2005) 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-77572005000200005 
787 0 |n https://doaj.org/toc/1678-7757 
787 0 |n https://doaj.org/toc/1678-7765 
856 4 1 |u https://doaj.org/article/60e4e6fa3f0147fc8b3ab70bf464c073  |z Connect to this object online.