Performance evaluation of the compounding robot, APOTECAchemo, for injectable anticancer drugs in a Japanese hospital

Abstract Background The accuracy, safety and feasibility of, the compounding robot APOTECAchemo were evaluated in the clinical practice of Japan. Methods Accuracy and precision of robotic preparations by APOTECAchemo was evaluated in 20 preparations of fluorouracil (FU) and cyclophosphamide (CPA) in...

Full description

Saved in:
Bibliographic Details
Main Authors: Takuya Iwamoto (Author), Takuya Morikawa (Author), Miki Hioki (Author), Hirofumi Sudo (Author), Demis Paolucci (Author), Masahiro Okuda (Author)
Format: Book
Published: BMC, 2017-04-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_677fdbce0eda4ffda5c3dbb8e86cfd9a
042 |a dc 
100 1 0 |a Takuya Iwamoto  |e author 
700 1 0 |a Takuya Morikawa  |e author 
700 1 0 |a Miki Hioki  |e author 
700 1 0 |a Hirofumi Sudo  |e author 
700 1 0 |a Demis Paolucci  |e author 
700 1 0 |a Masahiro Okuda  |e author 
245 0 0 |a Performance evaluation of the compounding robot, APOTECAchemo, for injectable anticancer drugs in a Japanese hospital 
260 |b BMC,   |c 2017-04-01T00:00:00Z. 
500 |a 10.1186/s40780-017-0081-z 
500 |a 2055-0294 
520 |a Abstract Background The accuracy, safety and feasibility of, the compounding robot APOTECAchemo were evaluated in the clinical practice of Japan. Methods Accuracy and precision of robotic preparations by APOTECAchemo was evaluated in 20 preparations of fluorouracil (FU) and cyclophosphamide (CPA) infusions by four pharmacists. Environmental and product contaminations with FU and CPA were evaluated by wipe testing. Robotic performance was compared with manual preparation in a biological safety cabinet. The number of robotic products, total compounding time and total pre-reconstitution time of lyophilized drugs between January 1, 2014 to December 31, 2015 were investigated. Results Robotic preparation resulted more accurate and precise (mean absolute dose error and coefficient of variation were 0.83 and 1.04% for FU and 0.52 and 0.59% for CPA) than those of manual preparation (respective values were 1.20 and 1.46% for FU and 1.70 and 2.20% for CPA). Drug residue was not detected from any of the prepared infusion bags with the robotic preparation, whereas FU was detected in two of four analyzed infusion bags with manual preparation. Average total time to make single anticancer drug preparation (compounding plus reconstitution of lyophilized drugs) was 6.11 min in the second half of 2015. During the study period, the highest percentage of production covered by APOTECAchemo was 70.4% of the total inpatient pharmacy activity. Conclusion Robotic preparation using APOTECAchemo should give substantial advantages in drug compounding for accuracy and safety and was able to be successfully worked in Mie university hospital. 
546 |a EN 
690 |a Robotic preparation 
690 |a Chemotherapy 
690 |a Wipe test 
690 |a APOTECAchemo 
690 |a Japanese hospital 
690 |a Pharmacy automation 
690 |a Therapeutics. Pharmacology 
690 |a RM1-950 
690 |a Pharmacy and materia medica 
690 |a RS1-441 
655 7 |a article  |2 local 
786 0 |n Journal of Pharmaceutical Health Care and Sciences, Vol 3, Iss 1, Pp 1-8 (2017) 
787 0 |n http://link.springer.com/article/10.1186/s40780-017-0081-z 
787 0 |n https://doaj.org/toc/2055-0294 
856 4 1 |u https://doaj.org/article/677fdbce0eda4ffda5c3dbb8e86cfd9a  |z Connect to this object online.