Effect of the sterilization on the cutting efficiency of ProTaper® and K3® endodontic instruments

Objective: This objective of this study was to evaluate the effect of sterilization on the cutting efficiency of rotary endodontic instruments, ProTaper® (Dentsply Maillefer Instruments S.A., Ballaigues, Suíça) and K3® (Sybron/Kerr, Orange, CA, USA), by the time X weighing method. Method: For this...

Full description

Saved in:
Bibliographic Details
Main Authors: Alex Niederauer Becker (Author), Graziele Borin (Author), Elias Pandonor Motcy de Oliveira (Author), Tiago André Fontoura de Melo (Author), Simone Soares Echeveste (Author)
Format: Book
Published: Faculdade São Leopoldo Mandic, 2009-12-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_67d603fb6fcf40e4979b47c40ff0b73f
042 |a dc 
100 1 0 |a Alex Niederauer Becker  |e author 
700 1 0 |a Graziele Borin  |e author 
700 1 0 |a Elias Pandonor Motcy de Oliveira  |e author 
700 1 0 |a Tiago André Fontoura de Melo  |e author 
700 1 0 |a Simone Soares Echeveste  |e author 
245 0 0 |a Effect of the sterilization on the cutting efficiency of ProTaper® and K3® endodontic instruments 
260 |b Faculdade São Leopoldo Mandic,   |c 2009-12-01T00:00:00Z. 
500 |a 0103-6971 
500 |a 1981-8637 
520 |a Objective: This objective of this study was to evaluate the effect of sterilization on the cutting efficiency of rotary endodontic instruments, ProTaper® (Dentsply Maillefer Instruments S.A., Ballaigues, Suíça) and K3® (Sybron/Kerr, Orange, CA, USA), by the time X weighing method. Method: For this purpose, 60 simulated canals and 66 instruments were used (36 of the ProTaper® system, (Dentsply Maillefer Instruments S.A., Ballaigues, Switzerland) and 30 of the K3® system (Sybron/Kerr, Orange, CA, USA). The rotary instruments were divided into four experimental groups. The instruments in Group A (K3®, Sybron/Kerr, Orange, CA, USA) and Group B (ProTaper®, Dentsply Maillefer Instruments S.A., Ballaigues, Switzerland), were subjected to a process of cleaning and sterilization. Whereas the instruments in Group C (K3®, Sybron/Kerr, Orange, CA, USA) and Group D (ProTaper®, Dentsply Maillefer Instruments S.A., Ballaigues, Switzerland), were subjected to the cleanness process only, and were not sterilized. Results: As regards analysis of the results, the Student's-t test was used, by which it was verified that the instruments in Group had greater cutting efficiency in comparison with those in Group B. Similarly the instruments in Group C had greater cutting efficiency in comparison with those in Group D, nevertheless, there was no statistically significant difference. Moreover, as regards the number times of use, there was also no statistically significant difference between the experimental groups. However in all the groups there was a decrease in the cutting efficiency with an increasing number of times of use. Conclusion: Thus, the cutting efficiency of rotary endodontic instruments K3® (Sybron/Kerr, Orange, CA, USA) and ProTaper® (Dentsply Maillefer Instruments S.A., Ballaigues, Suição) was not statistically affected after the use of 6 autoclave sterilization cycles. 
546 |a EN 
690 |a endodontics 
690 |a sterilization 
690 |a dental instruments 
690 |a endodontia 
690 |a esterilização 
690 |a instrumentos odontológicos 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n RGO: Revista Gaúcha de Odontologia, Vol 57, Iss 4, Pp 389-393 (2009) 
787 0 |n http://www.revistargo.com.br/include/getdoc.php?id=4223&article=1044&mode=pdf 
787 0 |n https://doaj.org/toc/0103-6971 
787 0 |n https://doaj.org/toc/1981-8637 
856 4 1 |u https://doaj.org/article/67d603fb6fcf40e4979b47c40ff0b73f  |z Connect to this object online.