Simulated damage of two implant debridement methods: Nonsurgical approach with Teflon and stainless steel hand scalers

Background: Mechanical scaling is the most common treatment of periodontal and peri-implant tissue infections. Aims: This study aimed to evaluate the effect of mechanical prophylactic therapy on the residual stresses in the implant and hand scaler. Settings and Design: For finite-element analysis, a...

Full description

Saved in:
Bibliographic Details
Main Authors: João Paulo Mendes Tribst (Author), Amanda Maria de Oliveira Dal Piva (Author), Dimas Renno de Lima (Author), Alexandre Luiz Souto Borges (Author), Marco Antonio Bottino (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2018-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_7a8f93b5c23443bbb0f8e6a9e43d6f09
042 |a dc 
100 1 0 |a João Paulo Mendes Tribst  |e author 
700 1 0 |a Amanda Maria de Oliveira Dal Piva  |e author 
700 1 0 |a Dimas Renno de Lima  |e author 
700 1 0 |a Alexandre Luiz Souto Borges  |e author 
700 1 0 |a Marco Antonio Bottino  |e author 
245 0 0 |a Simulated damage of two implant debridement methods: Nonsurgical approach with Teflon and stainless steel hand scalers 
260 |b Wolters Kluwer Medknow Publications,   |c 2018-01-01T00:00:00Z. 
500 |a 0972-124X 
500 |a 10.4103/jisp.jisp_201_18 
520 |a Background: Mechanical scaling is the most common treatment of periodontal and peri-implant tissue infections. Aims: This study aimed to evaluate the effect of mechanical prophylactic therapy on the residual stresses in the implant and hand scaler. Settings and Design: For finite-element analysis, an implant-supported prosthesis was created using modeling software with 3 mm of exposed threads. For simulation of a prophylactic mechanical debridement, the active face of the shank was disposed of in contact with the last thread exposed at a 90° angle. Materials and Methods: In the analysis software, the contacts were defined as rough between the instrument and the implant. The cortical bone was fixed and a load of 10 N was applied to the instrument cable. Two simulations were performed according to the instrument material: stainless steel or Teflon. Von-Mises results were obtained. Statistical Analysis Used: No statistical test was used, but, the 500 higher stress peaks in the implant and in the instrument were analyzed for qualitative comparison. Results: Mechanical prophylactic therapy generates higher residual stress on the implant with a stainless steel instrument. There was no difference between the materials for the active tip of the instrument, and the active portion of the shank was the region which concentrated more stress. Conclusions: It is suggested that hand scalers in Teflon are less damaging to the implant, but more susceptible to deformation and possible early failures. 
546 |a EN 
690 |a Cleaning instruments 
690 |a dental/oral implants 
690 |a hand instruments 
690 |a implant surfaces 
690 |a peri-implantitis 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Indian Society of Periodontology, Vol 22, Iss 4, Pp 340-344 (2018) 
787 0 |n http://www.jisponline.com/article.asp?issn=0972-124X;year=2018;volume=22;issue=4;spage=340;epage=344;aulast=Tribst 
787 0 |n https://doaj.org/toc/0972-124X 
856 4 1 |u https://doaj.org/article/7a8f93b5c23443bbb0f8e6a9e43d6f09  |z Connect to this object online.