Mobile Phone-Based Use of the Photoplethysmography Technique to Detect Atrial Fibrillation in Primary Care: Diagnostic Accuracy Study of the FibriCheck App

BackgroundMobile phone apps using photoplethysmography (PPG) technology through their built-in camera are becoming an attractive alternative for atrial fibrillation (AF) screening because of their low cost, convenience, and broad accessibility. However, some important questions concerning their diag...

Descrizione completa

Salvato in:
Dettagli Bibliografici
Autori principali: Proesmans, Tine (Autore), Mortelmans, Christophe (Autore), Van Haelst, Ruth (Autore), Verbrugge, Frederik (Autore), Vandervoort, Pieter (Autore), Vaes, Bert (Autore)
Natura: Libro
Pubblicazione: JMIR Publications, 2019-03-01T00:00:00Z.
Soggetti:
Accesso online:Connect to this object online.
Tags: Aggiungi Tag
Nessun Tag, puoi essere il primo ad aggiungerne!!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_82d2c75b2a7c48a3bf58bd756bfec3f0
042 |a dc 
100 1 0 |a Proesmans, Tine  |e author 
700 1 0 |a Mortelmans, Christophe  |e author 
700 1 0 |a Van Haelst, Ruth  |e author 
700 1 0 |a Verbrugge, Frederik  |e author 
700 1 0 |a Vandervoort, Pieter  |e author 
700 1 0 |a Vaes, Bert  |e author 
245 0 0 |a Mobile Phone-Based Use of the Photoplethysmography Technique to Detect Atrial Fibrillation in Primary Care: Diagnostic Accuracy Study of the FibriCheck App 
260 |b JMIR Publications,   |c 2019-03-01T00:00:00Z. 
500 |a 2291-5222 
500 |a 10.2196/12284 
520 |a BackgroundMobile phone apps using photoplethysmography (PPG) technology through their built-in camera are becoming an attractive alternative for atrial fibrillation (AF) screening because of their low cost, convenience, and broad accessibility. However, some important questions concerning their diagnostic accuracy remain to be answered. ObjectiveThis study tested the diagnostic accuracy of the FibriCheck AF algorithm for the detection of AF on the basis of mobile phone PPG and single-lead electrocardiography (ECG) signals. MethodsA convenience sample of patients aged 65 years and above, with or without a known history of AF, was recruited from 17 primary care facilities. Patients with an active pacemaker rhythm were excluded. A PPG signal was obtained with the rear camera of an iPhone 5S. Simultaneously, a single‑lead ECG was registered using a dermal patch with a wireless connection to the same mobile phone. PPG and single-lead ECG signals were analyzed using the FibriCheck AF algorithm. At the same time, a 12‑lead ECG was obtained and interpreted offline by independent cardiologists to determine the presence of AF. ResultsA total of 45.7% (102/223) subjects were having AF. PPG signal quality was sufficient for analysis in 93% and single‑lead ECG quality was sufficient in 94% of the participants. After removing insufficient quality measurements, the sensitivity and specificity were 96% (95% CI 89%-99%) and 97% (95% CI 91%-99%) for the PPG signal versus 95% (95% CI 88%-98%) and 97% (95% CI 91%-99%) for the single‑lead ECG, respectively. False-positive results were mainly because of premature ectopic beats. PPG and single‑lead ECG techniques yielded adequate signal quality in 196 subjects and a similar diagnosis in 98.0% (192/196) subjects. ConclusionsThe FibriCheck AF algorithm can accurately detect AF on the basis of mobile phone PPG and single-lead ECG signals in a primary care convenience sample. 
546 |a EN 
690 |a Information technology 
690 |a T58.5-58.64 
690 |a Public aspects of medicine 
690 |a RA1-1270 
655 7 |a article  |2 local 
786 0 |n JMIR mHealth and uHealth, Vol 7, Iss 3, p e12284 (2019) 
787 0 |n http://mhealth.jmir.org/2019/3/e12284/ 
787 0 |n https://doaj.org/toc/2291-5222 
856 4 1 |u https://doaj.org/article/82d2c75b2a7c48a3bf58bd756bfec3f0  |z Connect to this object online.