Post sterilization of intraoral scan body and the effect it has on the axes and distances between three adjacent implants: in-vitro study

Abstract Background The purpose of this pilot in-vitro study was to assess the effect of sterilization on the intra-implant axis, inter-implant axis, intra-implant distance and inter-implant distance of three implants in a straight line by using laboratory scanner (LBS) versus intra-oral scanner (IO...

Full description

Saved in:
Bibliographic Details
Main Authors: Gil Ben-Izhack (Author), Diva Lugassy (Author), Ofir Rosner (Author), Joseph Nissan (Author), Shir Rachmiel (Author), Yara Khoury (Author), Asaf Shely (Author)
Format: Book
Published: BMC, 2024-08-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_838fe4d68a1a4cf4b34b762c561cf0ce
042 |a dc 
100 1 0 |a Gil Ben-Izhack  |e author 
700 1 0 |a Diva Lugassy  |e author 
700 1 0 |a Ofir Rosner  |e author 
700 1 0 |a Joseph Nissan  |e author 
700 1 0 |a Shir Rachmiel  |e author 
700 1 0 |a Yara Khoury  |e author 
700 1 0 |a Asaf Shely  |e author 
245 0 0 |a Post sterilization of intraoral scan body and the effect it has on the axes and distances between three adjacent implants: in-vitro study 
260 |b BMC,   |c 2024-08-01T00:00:00Z. 
500 |a 10.1186/s12903-024-04664-1 
500 |a 1472-6831 
520 |a Abstract Background The purpose of this pilot in-vitro study was to assess the effect of sterilization on the intra-implant axis, inter-implant axis, intra-implant distance and inter-implant distance of three implants in a straight line by using laboratory scanner (LBS) versus intra-oral scanner (IOS) with intra-oral scan bodies (ISB). Methods: A printed 3D model with three internal hex analogs in the positions 15#,16#,17# was used. Zirkonzhan (ZZ) intra-oral scan body (ISB), two-piece titanium was used. The ZZ ISBs were scanned by 7 Series dental wings (LBS) and 30 times by Primescan (IOS) pre sterilization and 30 times post sterilization. For each scan (pre and post) stereolithography (STL) file was created and a comparison between all the scans pre sterilization and post sterilization were superimposed on the laboratory scan by using a 3D analyzing software. A Kolmogorov-Smirnov test performed followed by Wilcoxon Signed Ranks tests. (p < 0.05) Results: Post sterilization of the ZZ ISB, the mean errors were significantly increased for the inter-implant distances (p < 0.0005), intra-implant distances 1,2,3 (p < 0.0005), intra-implant axis 1,3 (p < 0.0005) and inter-implant axes 13,23 (p < 0.05). In contrast, the mean errors for intra-implant axis 2 (p < 0.0005) and inter-implant axis 12 (p < 0.0005) were significantly reduced. Conclusions: ZZ ISB showed changes in all four parameters after sterilization. The middle ISB had the largest changes in mean error regarding all four parameters. Sterilization process may affect the three-dimensional (3D) structure of the ZZ ISB after three cycles. There is a lack in the literature in this field and there is a need for further studies to explore the effect of sterilization (multiple cycles) on different ISBs and for creating an approved guidelines regarding the amount of sterilization for each ISB in the industry. 
546 |a EN 
690 |a CAD-CAM 
690 |a Scan abutment 
690 |a Intraoral scan body 
690 |a Sterilization 
690 |a Autoclave 
690 |a Implant axis 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n BMC Oral Health, Vol 24, Iss 1, Pp 1-9 (2024) 
787 0 |n https://doi.org/10.1186/s12903-024-04664-1 
787 0 |n https://doaj.org/toc/1472-6831 
856 4 1 |u https://doaj.org/article/838fe4d68a1a4cf4b34b762c561cf0ce  |z Connect to this object online.