Accuracy of probing attachment levels using a new computerized cemento-enamel junction probe

Background and Objectives: The assessment of clinical attachment level (CAL) represents the "gold standard" for diagnosing and monitoring periodontal disease. The aim of the present study was to evaluate the performance of the newly introduced cemento-enamel junction (CEJ) probe in detecti...

Full description

Saved in:
Bibliographic Details
Main Authors: R Deepa (Author), Shobha Prakash (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2012-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_8ffbb536de1448dd80c55a2bf338e69b
042 |a dc 
100 1 0 |a R Deepa  |e author 
700 1 0 |a Shobha Prakash  |e author 
245 0 0 |a Accuracy of probing attachment levels using a new computerized cemento-enamel junction probe 
260 |b Wolters Kluwer Medknow Publications,   |c 2012-01-01T00:00:00Z. 
500 |a 0972-124X 
500 |a 10.4103/0972-124X.94609 
520 |a Background and Objectives: The assessment of clinical attachment level (CAL) represents the "gold standard" for diagnosing and monitoring periodontal disease. The aim of the present study was to evaluate the performance of the newly introduced cemento-enamel junction (CEJ) probe in detecting CAL, using CEJ as a fixed reference point, and to compare the CEJ probe with the Florida stent probe (FSP) as well as with a standard manual probe, University of North Carolina-15 (UNC-15). Materials and Methods: Three examiners recorded the probing attachment level in 384 sites in case group (chronic periodontitis), and in 176 sites, in control group (healthy periodontal status), using the three probes. Subjects included both the sexes and ranged from 35 to 45 years. The experimental design was structured to balance the intra- and inter-examiner consistency at the same site during the two visits. Results: CEJ probe showed higher intra-and inter-examiner consistency over both FSP and UNC-15 in both the case and control groups. Frequency distribution of differences of various magnitudes of repeated measurements ≤1 mm was in the higher range of 86.8% to 87.5% for CEJ probe. The FSP was more reproducible than UNC-15 in detecting relative attachment level (RAL). Interpretation and Conclusion: CEJ automated probe was found to have greatest potential for accuracy and consistency in detecting CAL than FSP and UNC-15. The automated probes appeared to be more reproducible than manual probes. 
546 |a EN 
690 |a Attachment level 
690 |a cemento-enamel junction probe 
690 |a Florida stent probe 
690 |a manual probe 
690 |a periodontal disease 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Indian Society of Periodontology, Vol 16, Iss 1, Pp 74-79 (2012) 
787 0 |n http://www.jisponline.com/article.asp?issn=0972-124X;year=2012;volume=16;issue=1;spage=74;epage=79;aulast=Deepa 
787 0 |n https://doaj.org/toc/0972-124X 
856 4 1 |u https://doaj.org/article/8ffbb536de1448dd80c55a2bf338e69b  |z Connect to this object online.