Effect of room temperature and refrigerated storage on automated complete blood count: A longitudinal study

Objective: Delay in complete blood count (CBC) analysis is likely in laboratories with very high workload or when the samples are collected off site for testing in centralized laboratories. This research was done to study the stability of blood cell parameters in automated CBC at the room temperatur...

Full description

Saved in:
Bibliographic Details
Main Authors: Anu Jaya (Author), Naveen Kakkar (Author), MJoseph John (Author)
Format: Book
Published: Wolters Kluwer Medknow Publications, 2022-01-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_92c68f08576e43a4b26e301e33ea495f
042 |a dc 
100 1 0 |a Anu Jaya  |e author 
700 1 0 |a Naveen Kakkar  |e author 
700 1 0 |a MJoseph John  |e author 
245 0 0 |a Effect of room temperature and refrigerated storage on automated complete blood count: A longitudinal study 
260 |b Wolters Kluwer Medknow Publications,   |c 2022-01-01T00:00:00Z. 
500 |a 2348-3334 
500 |a 2348-506X 
500 |a 10.4103/cjhr.cjhr_140_20 
520 |a Objective: Delay in complete blood count (CBC) analysis is likely in laboratories with very high workload or when the samples are collected off site for testing in centralized laboratories. This research was done to study the stability of blood cell parameters in automated CBC at the room temperature (20°C-24°C) storage after 24 h and refrigerated (2°C-8°C) storage after 24 and 48 h. Materials and Methods: Automated CBC was done initially on 500 randomly selected blood samples in a 3-part hematology analyzer. Of these, 250 samples were stored at the room temperature (20°C-24°C), and the rest were refrigerated at 2°C-8°C. Re-analysis after 24 h was done for both groups. Data were analyzed using the Pearson correlation coefficient, ANOVA, and repeated-measure ANOVA. Results: Refrigerated samples stored at 2°C-8°C showed good stability for all hematological parameters (P > 0.5), except for mean platelet volume (MPV) in comparison with those at the room temperature storage. MPV showed a statistically significant (P < 0.001) rise on refrigerated storage. In the room temperature storage group, hematocrit (Hct), mean corpuscular volume, mean corpuscular hemoglobin concentration, red cell distribution width, MPV, and granulocyte percentage showed significant difference compared to the initial values. In the extended 48 h refrigerated storage group, all CBC parameters except Hct showed acceptable stability. Conclusion: The storage of blood samples at the room temperature and on refrigeration (2°C-8°C) for 24 h has shown acceptable stability of most of the automated CBC parameters. Refrigeration retained the true values over the room temperature storage and should be recommended if a delay in testing is expected. 
546 |a EN 
690 |a complete blood count 
690 |a refrigeration 
690 |a room temperature 
690 |a stability 
690 |a storage 
690 |a Medicine 
690 |a R 
690 |a Nursing 
690 |a RT1-120 
655 7 |a article  |2 local 
786 0 |n CHRISMED Journal of Health and Research, Vol 9, Iss 1, Pp 57-61 (2022) 
787 0 |n http://www.cjhr.org/article.asp?issn=2348-3334;year=2022;volume=9;issue=1;spage=57;epage=61;aulast=Jaya 
787 0 |n https://doaj.org/toc/2348-3334 
787 0 |n https://doaj.org/toc/2348-506X 
856 4 1 |u https://doaj.org/article/92c68f08576e43a4b26e301e33ea495f  |z Connect to this object online.