Investigation to Predict Primary Implant Stability Using Frictional Resistance Torque of Tap Drilling

Objectives: The purpose of this experimental study was to investigate the correlation between the frictional resistance torque of tap drilling prior to implant placement and the primary stability after implant placement. Material and Methods: Solid rigid polyurethane bone blocks of four different de...

Full description

Saved in:
Bibliographic Details
Main Authors: Kaien Wakamatsu (Author), Kazuya Doi (Author), Reiko Kobatake (Author), Yusuke Makihara (Author), Yoshifumi Oki (Author), Kazuhiro Tsuga (Author)
Format: Book
Published: Stilus Optimus, 2022-12-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_a5655ac8c12d40bba407916c0f0d4106
042 |a dc 
100 1 0 |a Kaien Wakamatsu  |e author 
700 1 0 |a Kazuya Doi  |e author 
700 1 0 |a Reiko Kobatake  |e author 
700 1 0 |a Yusuke Makihara  |e author 
700 1 0 |a Yoshifumi Oki  |e author 
700 1 0 |a Kazuhiro Tsuga  |e author 
245 0 0 |a Investigation to Predict Primary Implant Stability Using Frictional Resistance Torque of Tap Drilling 
260 |b Stilus Optimus,   |c 2022-12-01T00:00:00Z. 
500 |a 10.5037/jomr.2022.13401 
500 |a 2029-283X 
520 |a Objectives: The purpose of this experimental study was to investigate the correlation between the frictional resistance torque of tap drilling prior to implant placement and the primary stability after implant placement. Material and Methods: Solid rigid polyurethane bone blocks of four different densities were used in this study. A computerized surgical implant motor device was utilized to measure the frictional resistance torque of tap drilling. After the tap torque was measured, the dental implants were inserted at the prepared sites. During the implantation, the insertion torque was recorded, and resonance frequency analysis was performed, the value of which was calculated as the implant stability quotient. Thereafter, the correlation between the tap torque and the primary stability of the implant was evaluated and compared with the standard drilling protocol. Results: A significant positive correlation was found between the tap torque and insertion torque (Pearson's r = 0.88, P < 0.0001). Similarly, there was a positive correlation between the tap torque and implant stability quotient (Pearson's r = 0.69, P < 0.0001). Conclusions: These results suggest that measurement of the frictional resistance torque of tap drilling prior to implant placement could provide helpful information for implant primary stability. 
546 |a EN 
690 |a bone density 
690 |a dental implantation 
690 |a experimental implants 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n eJournal of Oral Maxillofacial Research, Vol 13, Iss 4, p e1 (2022) 
787 0 |n https://www.ejomr.org/JOMR/archives/2022/4/e1/v13n4e1ht.htm 
787 0 |n https://doaj.org/toc/2029-283X 
856 4 1 |u https://doaj.org/article/a5655ac8c12d40bba407916c0f0d4106  |z Connect to this object online.