The effect of different light-curing units on tensile strength and microhardness of a composite resin

The aim of this study was to evaluate the influence of different light-curing units on the tensile bond strength and microhardness of a composite resin (Filtek Z250 - 3M/ESPE). Conventional halogen (Curing Light 2500 - 3M/ESPE; CL) and two blue light emitting diode curing units (Ultraled - Dabi/Atla...

Full description

Saved in:
Bibliographic Details
Main Authors: Eduardo Batista Franco (Author), Patrícia Aleixo dos Santos (Author), Rafael Francisco Lia Mondelli (Author)
Format: Book
Published: University of São Paulo, 2007-12-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_a9570b57dac54b1b83d6c8a7b69eada7
042 |a dc 
100 1 0 |a Eduardo Batista Franco  |e author 
700 1 0 |a Patrícia Aleixo dos Santos  |e author 
700 1 0 |a Rafael Francisco Lia Mondelli  |e author 
245 0 0 |a The effect of different light-curing units on tensile strength and microhardness of a composite resin 
260 |b University of São Paulo,   |c 2007-12-01T00:00:00Z. 
500 |a 10.1590/S1678-77572007000600003 
500 |a 1678-7757 
500 |a 1678-7765 
520 |a The aim of this study was to evaluate the influence of different light-curing units on the tensile bond strength and microhardness of a composite resin (Filtek Z250 - 3M/ESPE). Conventional halogen (Curing Light 2500 - 3M/ESPE; CL) and two blue light emitting diode curing units (Ultraled - Dabi/Atlante; UL; Ultrablue IS - DMC; UB3 and UB6) were selected for this study. Different light intensities (670, 130, 300, and 600 mW/cm², respectively) and different curing times (20s, 40s and 60s) were evaluated. Knoop microhardness test was performed in the area corresponding to the fractured region of the specimen. A total of 12 groups (n=10) were established and the specimens were prepared using a stainless steel mold composed by two similar parts that contained a cone-shaped hole with two diameters (8.0 mm and 5.0 mm) and thickness of 1.0 mm. Next, the specimens were loaded in tensile strength until fracture in a universal testing machine at a crosshead speed of 0.5 mm/min and a 50 kg load cell. For the microhardness test, the same matrix was used to fabricate the specimens (12 groups; n=5). Microhardness was determined on the surfaces that were not exposed to the light source, using a Shimadzu HMV-2 Microhardness Tester at a static load of 50 g for 30 seconds. Data were analyzed statistically by two-way ANOVA and Tukey's test (p<0.05). Regarding the individual performance of the light-curing units, there was similarity in tensile strength with 20-s and 40-s exposure times and higher tensile strength when a 60-s light-activation time was used. Regarding microhardness, the halogen lamp had higher results when compared to the LED units. For all light-curing units, the variation of light-exposure time did not affect composite microhardness. However, lower irradiances needed longer light-activation times to produce similar effect as that obtained with high-irradiance light-curing sources. 
546 |a EN 
690 |a Photopolymerization 
690 |a LED 
690 |a Microhardness 
690 |a Tensile strength 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Applied Oral Science, Vol 15, Iss 6, Pp 470-474 (2007) 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-77572007000600003 
787 0 |n https://doaj.org/toc/1678-7757 
787 0 |n https://doaj.org/toc/1678-7765 
856 4 1 |u https://doaj.org/article/a9570b57dac54b1b83d6c8a7b69eada7  |z Connect to this object online.