Implant Diameter, Length, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks

The amount of energy necessary to place an implant in its seat, described as the integral of the torque-depth curve at insertion (I), has been validated as a reliable measure of primary stability. This study aimed to investigate whether (I) may detect the variations in primary stability caused by ch...

Full description

Saved in:
Bibliographic Details
Main Authors: Paolo Arosio (Author), Federico Arosio (Author), Danilo Alessio Di Stefano (Author)
Format: Book
Published: MDPI AG, 2020-06-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_add5994e111f42c5b1959440e6ca995b
042 |a dc 
100 1 0 |a Paolo Arosio  |e author 
700 1 0 |a Federico Arosio  |e author 
700 1 0 |a Danilo Alessio Di Stefano  |e author 
245 0 0 |a Implant Diameter, Length, and the Insertion Torque/Depth Integral: A Study Using Polyurethane Foam Blocks 
260 |b MDPI AG,   |c 2020-06-01T00:00:00Z. 
500 |a 10.3390/dj8020056 
500 |a 2304-6767 
520 |a The amount of energy necessary to place an implant in its seat, described as the integral of the torque-depth curve at insertion (I), has been validated as a reliable measure of primary stability. This study aimed to investigate whether (I) may detect the variations in primary stability caused by changes in the implant length or diameter better than the insertion torque (IT). Cylindric implants featuring a double-etched, sandblasted surface with different diameters or lengths were placed into monolithic polyurethane foam blocks with different densities that mimicked human bone. (I)-, (I)*-, IT-, IT*-diameter and -length plots ((I)* and IT* were the derived values corrected for undersizing) were drawn and the relation between (I), (I)*, IT, and IT* and the fixture diameter or length was investigated with correlation analysis. (I)* and IT* correlated better than (I) and IT with the fixture diameter; (I), (I)*, IT, and IT* correlated equally well with the fixture length. In all cases, the slopes of the lines best fitting the experimental data were greater for (I) or (I)* than IT or IT*, respectively. (I) or (I)* were better detectors than IT or IT* of the changes in primary stability that can be achieved by increasing the fixture diameter or length. 
546 |a EN 
690 |a primary stability 
690 |a torque-depth curve integral 
690 |a insertion torque 
690 |a undersizing 
690 |a fixture length 
690 |a fixture diameter 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Dentistry Journal, Vol 8, Iss 2, p 56 (2020) 
787 0 |n https://www.mdpi.com/2304-6767/8/2/56 
787 0 |n https://doaj.org/toc/2304-6767 
856 4 1 |u https://doaj.org/article/add5994e111f42c5b1959440e6ca995b  |z Connect to this object online.