Evaluation of temperature increase during in-office bleaching

ABSTRACT The use of light sources in the bleaching process reduces the time required and promotes satisfactory results. However, these light sources can cause an increase in the pulp temperature. Objective The purpose of the present study was to measure the increase in intrapulpal temperature induce...

Full description

Saved in:
Bibliographic Details
Main Authors: Rafael Francisco Lia MONDELLI (Author), Ana Flávia SOARES (Author), Eugenio Gabriel Kegler PANGRAZIO (Author), Linda WANG (Author), Sergio Kiyoshi ISHIKIRIAMA (Author), Juliana Fraga Soares BOMBONATTI (Author)
Format: Book
Published: University of São Paulo, 2016-04-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_b55abc5b14b14d9ab0cd87d2873e6be4
042 |a dc 
100 1 0 |a Rafael Francisco Lia MONDELLI  |e author 
700 1 0 |a Ana Flávia SOARES  |e author 
700 1 0 |a Eugenio Gabriel Kegler PANGRAZIO  |e author 
700 1 0 |a Linda WANG  |e author 
700 1 0 |a Sergio Kiyoshi ISHIKIRIAMA  |e author 
700 1 0 |a Juliana Fraga Soares BOMBONATTI  |e author 
245 0 0 |a Evaluation of temperature increase during in-office bleaching 
260 |b University of São Paulo,   |c 2016-04-01T00:00:00Z. 
500 |a 1678-7765 
500 |a 10.1590/1678-775720150154 
520 |a ABSTRACT The use of light sources in the bleaching process reduces the time required and promotes satisfactory results. However, these light sources can cause an increase in the pulp temperature. Objective The purpose of the present study was to measure the increase in intrapulpal temperature induced by different light-activated bleaching procedures with and without the use of a bleaching gel. Material and Methods A human maxillary central incisor was sectioned 2 mm below the cementoenamel junction. A K-type thermocouple probe was introduced into the pulp chamber. A 35% hydrogen peroxide bleaching gel was applied to the vestibular tooth surface. The light units used were a conventional halogen, a hybrid light (only LED and LED/Laser), a high intensity LED, and a green LED light. Temperature increase values were compared by two-way ANOVA and Tukey´s tests (p<0.05). Results There were statistically significant differences in temperature increases between the different light sources used and between the same light sources with and without the use of a bleaching gel. The presence of a bleaching gel generated an increase in intra-pulpal temperature in groups activated with halogen light, hybrid light, and high intensity LED. Compared to the other light sources, the conventional halogen lamp applied over the bleaching gel induced a significant increase in temperature (3.83±0.41°C). The green LED unit with and without gel application did not produce any significant intrapulpal temperature variations. Conclusion In the present study, the conventional halogen lamp caused the highest increase in intrapulpal temperature, and the green LED caused the least. There was an increase in temperature with all lights tested and the maximum temperature remained below the critical level (5.5°C). The addition of a bleaching gel led to a higher increase in intrapulpal temperatures. 
546 |a EN 
690 |a Tooth bleaching 
690 |a Light 
690 |a Temperature 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Applied Oral Science, Vol 24, Iss 2, Pp 136-141 (2016) 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-77572016000200136&lng=en&tlng=en 
787 0 |n https://doaj.org/toc/1678-7765 
856 4 1 |u https://doaj.org/article/b55abc5b14b14d9ab0cd87d2873e6be4  |z Connect to this object online.