Effect of pontic framework design on the fracture resistance of implant-supported all-ceramic fixed partial dentures

OBJECTIVE: The purpose of this study was to compare the fracture resistance of implant-supported all-ceramic fixed partial dentures, which have three different pontic designs. MATERIAL AND METHODS: Two implants were placed in a metal model simulating mandibular left second premolar and mandibular le...

Full description

Saved in:
Bibliographic Details
Main Authors: Ozgur Inan (Author), Asli Secilmis (Author), Oguz Eraslan (Author)
Format: Book
Published: University of São Paulo, 2009-10-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_bf2ff1dc9d1a4e64aae8bc7a6a9d6ea2
042 |a dc 
100 1 0 |a Ozgur Inan  |e author 
700 1 0 |a Asli Secilmis  |e author 
700 1 0 |a Oguz Eraslan  |e author 
245 0 0 |a Effect of pontic framework design on the fracture resistance of implant-supported all-ceramic fixed partial dentures 
260 |b University of São Paulo,   |c 2009-10-01T00:00:00Z. 
500 |a 10.1590/S1678-77572009000500032 
500 |a 1678-7757 
500 |a 1678-7765 
520 |a OBJECTIVE: The purpose of this study was to compare the fracture resistance of implant-supported all-ceramic fixed partial dentures, which have three different pontic designs. MATERIAL AND METHODS: Two implants were placed in a metal model simulating mandibular left second premolar and mandibular left second molar. Thirty standardized 3-unit all-ceramic fixed partial dentures with biconvex, convex or concave pontic designs were fabricated using IPS e.max system (n=10). Afterwards, specimens were centrally loaded on the pontics until failure with a universal testing machine. Results were analyzed by Kruskal-Wallis and Mann-Whitney U tests at 5% significance level. RESULTS: The fracture resistance values of all-ceramic fixed partial dentures designed with biconvex, convex or concave pontics were 349.71, 438.20 and 300.78 N, respectively. There were no statistically significant differences between the fracture resistances of the groups (p>0.05), except for convex and concave groups (p<0.05 and p=0.009, respectively). CONCLUSIONS: Convex design showed the best mechanical properties as demonstrated by the high values of fracture resistance. 
546 |a EN 
690 |a Pontic 
690 |a Design 
690 |a Dental implants 
690 |a Fixed prosthesis 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Journal of Applied Oral Science, Vol 17, Iss 5, Pp 533-538 (2009) 
787 0 |n http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1678-77572009000500032 
787 0 |n https://doaj.org/toc/1678-7757 
787 0 |n https://doaj.org/toc/1678-7765 
856 4 1 |u https://doaj.org/article/bf2ff1dc9d1a4e64aae8bc7a6a9d6ea2  |z Connect to this object online.