Normal and reverse flow injectionâspectrophotometric determination of thiamine hydrochloride in pharmaceutical preparations using diazotized metoclopramide

Simple and sensitive normal and reverse flow injection methods for spectrophotometric determination of thiamine hydrochloride (THC) at the microgram level were proposed and optimized. Both methods are based on the reaction between THC and diazotized metoclopramide in alkaline medium. Beerâs law was...

Full description

Saved in:
Bibliographic Details
Main Authors: Mouayed Q. Al Abachi (Author), Hind Hadi (Author)
Format: Book
Published: Elsevier, 2012-10-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Simple and sensitive normal and reverse flow injection methods for spectrophotometric determination of thiamine hydrochloride (THC) at the microgram level were proposed and optimized. Both methods are based on the reaction between THC and diazotized metoclopramide in alkaline medium. Beerâs law was obeyed over the range of 10â300 and 2â90 μg/mL, the limits of detection were 2.118 and 0.839 μg/mL and the sampling rates were 80 and 95 injections per hour for normal and reverse flow injection methods respectively. The application of both methods to commercially available pharmaceuticals produced acceptable results. The flow system is suitable for application in quality control processes. Keywords: Thiamine, Metoclopramide, Flow injection, Spectrophotometry
Item Description:2095-1779
10.1016/j.jpha.2012.04.005