Cyclic fatigue resistance of different continuous rotation and reciprocating endodontic systems
Objective: The objective of this study is to compare the cyclic fatigue resistance of nine types of endodontic instruments of nickel-titanium. Materials and Methods: Five files of 25 mm of length of each group: Reciproc (RC) R25; WaveOne (WO) Primary; Unicone (UC) L25 25/0.06; K3XF 25/0.06; ProTaper...
Saved in:
Main Authors: | , , , , , , |
---|---|
Format: | Book |
Published: |
Wolters Kluwer Medknow Publications,
2016-01-01T00:00:00Z.
|
Subjects: | |
Online Access: | Connect to this object online. |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objective: The objective of this study is to compare the cyclic fatigue resistance of nine types of endodontic instruments of nickel-titanium. Materials and Methods: Five files of 25 mm of length of each group: Reciproc (RC) R25; WaveOne (WO) Primary; Unicone (UC) L25 25/0.06; K3XF 25/0.06; ProTaper Universal F2 (PTF2); ProTaper Next X2 (PTX2); Mtwo 25/0.06; BioRaCe 25/0.06; One Shape L25 25/0.06 were subjected to a cyclic fatigue resistance test on a mechanical apparatus. The mean fracture time was analyzed statistically by one-way analysis of variance and Tukey's honest significant difference post hoc test, with significance set at P< 0.05. Conclusion: It was observed that the groups PTX2, RC, R25, UC L25 25/0.06, and WO Primary presented greater cyclic fracture resistance than the other groups (P< 0.001). |
---|---|