Comparative study of the accuracy of CBCT implant site measurements using different software programs

Purpose: To measure and compare the accuracy of the linear dimensions of implant sites recorded from cone beam computed tomography (CBCT) images using Blue Sky Plan, coDiagnostiX, and RadiAnt. Materials and Methods: Five human dry skulls were imaged with a CBCT device then sectioned to obtain sample...

Full description

Saved in:
Bibliographic Details
Main Author: Asma'a A. Al-Ekrish (Author)
Format: Book
Published: Elsevier, 2021-09-01T00:00:00Z.
Subjects:
Online Access:Connect to this object online.
Tags: Add Tag
No Tags, Be the first to tag this record!

MARC

LEADER 00000 am a22000003u 4500
001 doaj_eea7f39a84d14c2c8c6675c9fc3bd1e9
042 |a dc 
100 1 0 |a Asma'a A. Al-Ekrish  |e author 
245 0 0 |a Comparative study of the accuracy of CBCT implant site measurements using different software programs 
260 |b Elsevier,   |c 2021-09-01T00:00:00Z. 
500 |a 1013-9052 
500 |a 10.1016/j.sdentj.2020.07.003 
520 |a Purpose: To measure and compare the accuracy of the linear dimensions of implant sites recorded from cone beam computed tomography (CBCT) images using Blue Sky Plan, coDiagnostiX, and RadiAnt. Materials and Methods: Five human dry skulls were imaged with a CBCT device then sectioned to obtain sample transverse cross-sections of the edentulous ridges, and the height and width of the ridge were measured with a digital caliper to provide the gold standard measurements. The CBCT datasets were exported in DICOM format and imported into the three test software programs which were used to obtain reformatted sectional images corresponding to the sample transverse cross-sections, and the height and width of the edentulous ridge was recorded using the linear measurement tool. Reliability of the measurements were measured using the intraclass correlation coefficient. One-sample t-test (test value: zero) was used to test the statistical significance of the mean of the absolute errors for each software program. Analysis of Variance with Repeated Measures was used to test the statistical significance of the difference between the means of the absolute errors obtained by the different software programs. Statistical significance was set at a p-value of 0.05. Results: The reliability of the gold standard and image measurements were excellent. All three software programs demonstrated a statistically significant mean absolute measurement error of between 0.43 and 0.56 mm (p-value < 0.01), but no significant difference in error values was found between any of the tested programs (p- value = 0.18). Conclusions: There was no statistically significant difference in accuracy of linear CBCT measurements of implant sites recorded using Blue Sky Plan, coDiagnostiX, and RadiAnt. 
546 |a EN 
690 |a Diagnostic Imaging 
690 |a Dimensional measurement accuracy 
690 |a Cone beam computed tomography 
690 |a Dental implants 
690 |a Software Validation 
690 |a CBCT 
690 |a Medicine 
690 |a R 
690 |a Dentistry 
690 |a RK1-715 
655 7 |a article  |2 local 
786 0 |n Saudi Dental Journal, Vol 33, Iss 6, Pp 355-361 (2021) 
787 0 |n http://www.sciencedirect.com/science/article/pii/S1013905220307446 
787 0 |n https://doaj.org/toc/1013-9052 
856 4 1 |u https://doaj.org/article/eea7f39a84d14c2c8c6675c9fc3bd1e9  |z Connect to this object online.