Comparison between the HemoCue® and an automated counter for measuring hemoglobin Comparação entre o HemoCue® e um contador automático para determinação de hemoglobina
To determine the precision and agreement of the hemoglobin (Hb) measurements in capillary and venous blood samples by the HemoCue® and an automated counter. Hb was determined by both equipaments in blood samples of 29 pregnant women. The HemoCue® showed low repeatability of Hb measurements in duplic...
Saved in:
Main Authors: | , , , |
---|---|
Format: | Book |
Published: |
Universidade de São Paulo,
2004-08-01T00:00:00Z.
|
Subjects: | |
Online Access: | Connect to this object online. |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To determine the precision and agreement of the hemoglobin (Hb) measurements in capillary and venous blood samples by the HemoCue® and an automated counter. Hb was determined by both equipaments in blood samples of 29 pregnant women. The HemoCue® showed low repeatability of Hb measurements in duplicate in capillary (CR=0.53 g/dL, CV=13.6%) and venous blood (CR=0.53 g/dL, CV=13.6%). Hb measurements in capillary blood were higher than those in venous blood (12.4 and 11.7 g/dL, respectively; p<0.05). There was high agreement between Hb in capillary blood by the HemoCue® and in venous blood by the counter (r icc=0.86; p<0.01), and also between the diagnosis of anemia by both equipments (k=0.81; p<0.01). The HemoCue® seems to be more appropriate for capillary blood and require training of the measurers.<br>Com o objetivo de avaliar a precisão e concordância na determinação de hemoglobina (Hb) em sangue capilar e venoso, utilizando-se o HemoCue® e um contador automático, foram estudadas 29 gestantes. O HemoCue® apresentou baixa repetibilidade para as medidas em duplicata em sangue capilar (CR=0,53 g/dL; CV=13,6%) e venoso (CR=0,53 g/dL; CV=13,6%). As medidas de Hb obtidas em sangue capilar foram mais altas do que as obtidas em sangue venoso (12,4 e 11,7 g/dL, respectivamente; p<0,05). Verificou-se alta concordância entre medidas de Hb em sangue capilar obtidas pelo HemoCue® e em sangue venoso pelo contador (r icc=0,86; p<0,01), bem como entre os diagnósticos de anemia realizados nos dois equipamentos (k=0,81; p<0,01). O HemoCue® parece ser mais apropriado para análise de Hb em sangue capilar, mas é necessário um treinamento dos analisadores. |
---|---|
Item Description: | 10.1590/S0034-89102004000400017 0034-8910 1518-8787 |