Timing of Annual Output Calibration of Radiotherapy Linear Accelerators
<p>The output of radiotherapy Linear Accelerators (Linacs) is calibrated following national or international protocols like TG-51 by the American Association of Physicists in Medicine [1] or TRS-398 by the International Atomic Energy Agency [2], annually on or before the annivers...
Saved in:
Main Author: | |
---|---|
Format: | Book |
Published: |
International Journal of Radiology and Radiation Oncology - Peertechz Publications,
2016-01-23.
|
Subjects: | |
Online Access: | Connect to this object online. |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | <p>The output of radiotherapy Linear Accelerators (Linacs) is calibrated following national or international protocols like TG-51 by the American Association of Physicists in Medicine [1] or TRS-398 by the International Atomic Energy Agency [2], annually on or before the anniversary date of the commissioning of each Linac.The anniversary dates fall at random times around the year. Is there a better time for annual output calibration than on random anniversary dates? In view of the recent report of seasonal output variation [3,4],it is time to consider performing annual output calibration in the months not having extreme air humidity.</p> |
---|---|