Iatrogenic third-degree burn caused by off-label use of an infrared radiant heat lamp in a patient with accidental hypothermia

We report a case of third-degree burn involving 5% of the total body surface area caused by off-label use of Hanautherm®, an infrared radiant heat lamp for bed-warming, in a patient with severe accidental hypothermia. This iatrogenic burn required repeated surgical interventions and long-term hospit...